Filter Test Principle
Apr 15, 2022| ※ Test principle of bubble point method: when the filter membrane and filter element are completely soaked with a certain solution, and then pressurized on one side by the air source (there is an air intake control system in our instrument, which can stabilize the pressure and adjust the air intake), with the When the pressure increases, the gas is released from one side of the filter membrane, showing that bubbles of different sizes and numbers appear on one side of the membrane. The corresponding pressure value is judged by the instrument to be the bubble point.
※ Test principle of diffusion flow method: Diffusion flow test means that when the gas pressure is 80% of the bubble point value of the filter element, there is not a large amount of gas perforation, but a small amount of gas first dissolves into the liquid phase diaphragm. , and then diffuse from the liquid phase into the gas phase on the other side, this part of the gas is called diffusion flow.
※ Why is the diffusion flow method better: The bubble point value is only a qualitative value, and it is a long process from the beginning of bubbling to the last group bubbling, which cannot be accurately quantified. The measurement of the diffusion flow value is a quantitative value, which can not only accurately determine the integrity of the filter, but also reflect the porosity, flow rate and effective filtration area of the membrane. This is why foreign manufacturers use diffusion Reasons for flow test integrity.
※ Test principle of water intrusion method: The water intrusion method is specially used for the test of hydrophobic filter element. The hydrophobic membrane resists water, and the smaller the pore size, the greater the pressure required to squeeze water into the hydrophobic membrane. Therefore, under a certain pressure, the flow rate of water squeezed into the filter membrane is measured to determine the pore size of the filter element.